Abstract

An atomic absorption spectrophotometric (AAS) method applicable to a wide variety of mercury-containing drugs has been developed and subjected to a collaborative study. Samples were digested with a water-HCl-HNO3 (4 + 3 + 1) mixture, and the mercury was determined in solution by AAS. High levels of mercury were measured with a conventional air-acetylene flame, whereas low levels were measured by the flameless technique. Each of 7 collaborators received duplicate synthetic samples of a tincture, an ophthalmic solution, and an antiseptic solution, and duplicate commercial samples of an ointment and an injectable. The overall mean value found by collaborators for mercury in these samples was 100.13%. The corresponding overall repeatability SD (CV, %) and reproducibility SD (CV, %) values were 2.18 (2.18) and 3.38 (3.38), respectively. The proposed AAS method has been adopted official first action.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.