Abstract

In this paper direct determination of silver collected on ion-exchange filter paper was studied. 50 ml of sample solution of which pH was adjusted to 3.3 with hydrochloric acid was passed through a sheet of ion-exchange filter paper, and after the filter paper was dried, it was punched with a 2.5 mm puncher. The small tablet was put onto a tungsten atomizing furnace and electrothermally heated. The calibration curve was linear at Ag+ concentrations below 0.3 ng/ml, and 1% absorption sensitivity was 1.2×10-3 ng/ml. Coexsisting 104-fold Na+, 103-fold Cu2+, and 103-fold Pb2+ did not affect the absorbance of Ag. Detection limit of this method is considered to lower by increasing the volume of sample solution.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call