Abstract
A three-coil tungsten wire is used as electrode for the preconcentration of cadmium, which is then placed in a graphite tube for atomization and atomic absorption measurement. The heating parameters of the graphite furnace are optimized using the Modified and Weighted Centroid Simplex Method (MWCS), and computer program for automatic calculation. Sulphuric acid is selected as the supporting electrolyte for electrodeposition. The linear range of the calibration graph is 0.00–0.55 ng/ml. The detection limit is 0.01 ng/ml. For 0.1 ng/ml cadmium the coefficient of variation is 3.35% for ten parallel determinations. No interference occurs in the presence of more than 20 coexisting ions. Traces of cadmium in urine of normal people and in river water and the recoveries for cadmium are determined. The results are satisfactory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.