Abstract

When a tissue slice pretreated with gold nanoparticles is irradiated with a focused 532-nm continuous wave laser, desorption is observed to be uniform and stable, and its shape is sharp and steep. However, since the desorbed molecules are still electrically neutral particles in atmospheric pressure (AP), additional procedure of ionizing them is necessary for AP mass spectrometry (AP-MS) analysis. Therefore, the authors have combined an electrospray device with a simple chamber connected to the airflow-assisted particle transport equipment mounted at the mass spectrometer inlet. Subsequent ionization processes using an electrospray device enable the detection of several types of diacylglycerol molecules above 500 Da, which cannot be detected with the use of AP plasma jets. The authors also developed a remote AP-MS using a long and flexible sampling probe and a fiber laser with a slight modification of the proposed AP desorption and ionization method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.