Abstract

We report an atmospheric pressure (AP) matrix-assisted laser desorption/ionization (MALDI) mass spectrometry imaging (MSI) setup with a lateral resolution of 1.4 μm, a mass resolution greater than 100,000, and accuracy below ±2 p.p.m. We achieved this by coupling a focusing objective with a numerical aperture (NA) of 0.9 at 337 nm and a free working distance of 18 mm in coaxial geometry to an orbitrap mass spectrometer and optimizing the matrix application. We demonstrate improvement in image contrast, lateral resolution, and ion yield per unit area compared with a state-of-the-art commercial MSI source. We show that our setup can be used to detect metabolites, lipids, and small peptides, as well as to perform tandem MS experiments with 1.5-μm2 sampling areas. To showcase these capabilities, we identified subcellular lipid, metabolite, and peptide distributions that differentiate, for example, cilia and oral groove in Paramecium caudatum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.