Abstract

A device utilizing atmospheric pressure glow discharge as the second excitation source coupled with laser ablation (LA) for direct solid sampling was developed, with few operating costs and low gas consumption. This new device was first utilized for the highly sensitive determination of Zn, Pb, and Cd elements in complex matrix soil samples. It also provided a new method for monitoring these three trace elements in soil samples. Good linearity was observed in the quantitative results for Zn, Pb, and Cd detection, and the respective linear correlation coefficients (R2) were 0.9953, 0.9897, and 0.9961. Moreover, the limit of detection (LOD) of 0.68, 2.71, and 0.31 mg kg−1 were achieved for Zn, Pb, and Cd, respectively; the LOD of Zn reduced by more than one order of magnitude compared to that observed in laser-induced breakdown spectroscopy results. In addition, the quantitative analysis results showed good agreement with the certified values and those obtained of ICP optical emission spectrometry, proving the detection accuracy and practicability of the developed device.