Abstract

We report a case of a new syndrome, late asymmetric vault of the AT45 Crystalens intraocular lens (IOL) (Eyeonics Inc.). The late vault, which occurred in 6 month after implantation of the IOL, caused a decrease in uncorrected distance and near acuities and minus-cylinder axis perpendicular to the haptics. Treatment with neodymium:YAG laser capsulotomy and fibrotic band release was successful.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.