Abstract

AbstractThe asymmetric calibration method is a very interesting technology for differential output probes. It has been proven that the asymmetric calibration method (ACM) can broaden the application scenarios of the electromagnetic field dual probe. However, ACM is not verified on a back‐to‐back double‐loop differential magnetic field probing system. This paper calibrates a back‐to‐back double‐loop differential magnetic field probing system by inserting a connector for creating an asymmetry using ACM. The calibration results show that ACM can be used to calibrate an asymmetric back‐to‐back double‐loop differential magnetic field probing system. The calibration results are further verified by measuring the standing wave magnetic field, and the verification results show that the asymmetric calibration method is effective in eliminating the asymmetry of the back‐to‐back double‐loop differential magnetic field probing system and the work frequency band reaches up to 12 GHz.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call