In many monochromatic x-ray applications, it is important to maintain a constant direction of the monochromatized beam, preferably parallel to the direction of the original beam. For the most exacting applications, independent-crystal two-bounce monochromators can be used. These however, are costly and cumbersome. Therefore, in less demanding situations or when space is at premium, two-bounce channel-cut monochromators are commonly used. Unfortunately, while the direction of the beam emerging from a channel-cut monochromator is fixed, the beam offset varies with the angle of incidence, i.e., with the energy of the monochromatic beam. This complicates experiments in which it is important to maintain the beam on a fixed spot of the sample. We show, however, that the variation of beam offset in channel-cut monochromators can be greatly reduced by introducing a small amount of asymmetry into the cut. The exact amount of asymmetry (typically a few degrees) depends, of course, on the crystal and the energy range. We will present a full mathematical analysis and a design of a working monochromator based on this scheme.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.