A heterodyne receiver making use of a SIS waveguide mixer with integrated horn and single backshort tuner has been built. It has been used for a series of airborne measurements of atmospheric trace gases, such as HCl and ClO, above northern Europe during the winter of 1993 and 1994. The receiver is suitable for measurements in the range of 625-760 GHz and shows stable operation in the airplane together with high sensitivity. Best achieved noise temperatures are T/sub DSB/=310 K at 708 GHz in the laboratory and T/sub SSB/=1500 K at 625 GHz for the complete system in the airplane. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.