Abstract
Experimental: and theoretical studies were conducted to investigate low and non-linear responses in sub-micro molar-level suppressed ion chromatography with a hydroxide eluent. A calculated response was derived using experimentally determined detector effluent ion composition data and compared with measured experimental responses. The calibration curve was non-linear, and its slope varied considerably with different instrumental setups and operating conditions. The non-linearity of the solution conductivity response was determined by two acid-base equilibria of water and carbonic acid, and fluoride ion. By using eluent contaminated with para-toluene sulfonate at micro molar level, the non-linear response was greatly alleviated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.