Abstract

Resonance frequency analysis (RFA) has been introduced as a noninvasive method to clinically estimate the stability of dental implant systems. The objective of this study was to examine whether implant stability quotient (ISQ) values of RFA can account for mechanical stability of the dental implant system, which is assessed using dynamic mechanical analysis (DMA). Fifty-seven screw-type titanium dental implants were placed in artificial polyurethane foams with seven different thicknesses (3.5 to 12 mm) and eight edentulous mandibles of human cadavers (four men and four women, 79.11 ± 13.48 years). After the ISQ values, insertion torque, and static stiffness of each implant system were measured, the DMA was performed to assess dynamic stiffness and viscoelastic tan δ. The ISQ value had strong positive correlations with thickness, insertion torque, static and dynamic stiffness, and a negative correlation with tan δ of implant systems in artificial bone blocks (r = 0.769 to 0.992, p < .043). However, the ISQ value was correlated with only the insertion torque of implant systems in human mandibles (p < .049). The ISQ values could reflect mechanical stability of the dental implant system under the controlled condition of homogeneous density in simple dimensions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.