Abstract
Quantitative ultrasound (QUS) methods have been shown to be useful in the assessment of bone status. Nevertheless, ultrasound transmission depends on a variety of skeletal parameters, and a detailed understanding of ultrasound propagation through bone is important for the accurate interpretation of QUS results. We determined correlations between morphometrical and QUS parameters at the shaft of human finger phalanges using ultrasound transmission under different angles. 54 phalanges were measured in vitro using QUS and micro computed tomography (/spl mu/-CT). /spl mu/-CT was used for the derivation of geometrical parameters and cortical porosity. Significant and in part independent correlations between QUS and morphometry could be found indicating that QUS methods might be able to estimate different aspects of bone quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.