Abstract
1 Inje University Haeundae Paik Hospital, Busan, Korea Background: Obesity is associated with various chronic diseases, especially abdominal fat affect cardiovascular disease, metabolic disease, diabetes mellitus and mortality. The aim of this study was to estimate the accuracy of the visceral fat area measured by bioelectric impedence analysis (BIA) that easily used in many clinical settings, and to compare with the visceral fat area measured by CT and other abdominal obesity measures. Methods: Participants were 71 premenopausal adult women whose BMI was over 23 kg/m 2 . BMI, waist circumference, total abdominal fat area, visceral fat area, subcutaneous fat area by CT and truncal fat by dual-energy X-ray absorptiometry (DEXA) were measured. Visceral fat area measured by Inbody 720 were compared with variables examining abdominal obesity using partial correlation analysis and Bland-Altman analysis. The accuracy of the visceral obesity by BIA was compared with that diagnosed by CT as gold standard. Results: There was significant difference between visceral fat area measured by CT and BIA in women below BMI 30
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.