Abstract

BackgroundProgressive keratoconus can lead to severely impaired vision, but there is currently no consensus on the definition of progressive disease. Errors in the measurement of the parameters commonly used to establish progressive disease were evaluated in an attempt to determine the limits at which a true change in the values can be detected. The possible association between measurement error and disease severity was also investigated to evaluate the need for limits based on disease severity.MethodsSixty-one eyes were studied in 61 patients with keratoconus. Four replicate measurements were made in each patient using a Scheimpflug-based tomographic system (denoted the PC) and an auto-keratometer (denoted the AK). The repeatability coefficient, i.e., the level below which differences between two measurements are found in 95% of paired observations, was calculated. Patients were further divided into three groups based on disease severity (parameter magnitude).ResultsIncreasing magnitude of all the keratometric parameters investigated was significantly associated with increasing measurement errors, and thus worse repeatability. The maximum keratometry value (Kmax) was the least repeatable parameter (1.23 D, 95% CI 1.11–1.35 D) and showed the strongest association between parameter magnitude and measurement error. The repeatability coefficient ranged between 0.32 and 1.62 D, depending on disease severity. The most repeatable parameter was the flattest central keratometry value (K1), measured with the PC (0.51 D, 95% CI 0.46–0.56 D) and the AK (0.54 D, 95% CI 0.48–0.59 D). K1 showed the weakest association between parameter magnitude and measurement error. The repeatability coefficient for K1 ranged between 0.40 and 0.54 D when using the PC, and between 0.34 and 0.70 D when using the AK in the three groups.ConclusionsThe association between the magnitude of the keratometric parameters and their measurement errors suggests that limits should be based on disease severity to ensure reliable detection of progressive keratoconus. Further studies are, however, required.

Highlights

  • Keratoconus is a corneal disease that can lead to severely impaired vision

  • Increasing magnitude of all the keratometric parameters investigated was significantly associated with increasing measurement errors, and worse repeatability

  • The maximum keratometry value (Kmax) was the least repeatable parameter (1.23 D, 95% CI 1.11–1.35 D) and showed the strongest association between parameter magnitude and measurement error

Read more

Summary

Introduction

Keratoconus is a corneal disease that can lead to severely impaired vision. The instruments currently available for corneal imaging allow early detection of keratoconus [8], but there is no consensus on the definition of progressive keratoconus, which is the common indication for CXL. An increasing measurement error with increasing disease severity could explain some of the differences reported in repeatability, necessitating appropriate methods of calculating repeatability [12]. Progressive keratoconus can lead to severely impaired vision, but there is currently no consensus on the definition of progressive disease. Errors in the measurement of the parameters commonly used to establish progressive disease were evaluated in an attempt to determine the limits at which a true change in the values can be detected. The possible association between measurement error and disease severity was investigated to evaluate the need for limits based on disease severity

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.