Abstract
Bleeding is the main cause of postoperative complications during hepatic surgery. Blood loss and transfusions increase tumor recurrence in liver metastases from colorectal cancer. A high intensity focused ultrasound (HIFU) device with an integrated ultrasound imaging probe was developed for the treatment of colorectal liver metastasis. The HIFU toroidal-shaped transducer contains 256 elements (working frequency: 3 MHz) and can create a single conical lesion of 7 cm3 in 40 s. Then, the volume of treatment can be significantly increased by juxtaposing single lesions. Presented here is the use of this device in an animal model as a complementary tool to improve surgical resection in the liver. Before transecting the liver, a wall of coagulative necrosis was performed using this device in order to minimize blood loss and dissection time during hepatectomy. Resection assisted by HIFU was compared to classical dissections with clamping [intermittent Pringle maneuver (IPM) group] and without clamping (control group). For each technique, 14 partial liver resections were performed in seven pigs. Blood loss per dissection surface area and resection time were the main outcome parameters. Conserving liver blood inflow during hepatic resection assisted by HIFU did not increase total blood loss (7.4 +/- 3.3 ml cm(-2)) compared to hepatic resection performed during IPM and controlled blood inflow (11.2 +/- 2.2 ml cm(-2)). Lower blood loss was measured on average when using HIFU, even though difference with clamping (IPM) was not statistically significant (p = 0.09). Resection assisted by HIFU reduced blood loss by 50% compared to control group (14.0 +/- 3.4 ml cm(-2), p = 0.03). The duration of transection when using HIFU (13 +/- 3 min) was significantly lower compared to clamping (23 +/- 4 min, p < 0.01) and control (18 +/- 3 min, p = 0.02). Precoagulation also resulted in sealing blood vessels with a diameter of less than 5 mm, and therefore the number of clips needed in the HIFU group was significantly lower (0.8 +/- 0.2 cm(-2)) when compared to clamping (1.6 +/- 0.2 cm(-2), p < 0.01) and control (1.8 +/- 0.4 cm(-2), p < 0.01). This method holds promise for future clinical applications in resection of liver metastases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.