Abstract

Deep Inspiration Breath-Hold (DIBH) techniques for breast cancer radiation therapy (RT) have reduced cardiac dose compared to Free Breathing (FB). Recently, a voluntary deep inspiration breath-hold (vDIBH) technique was established using in-room lasers and skin tattoos to monitor breath-hold. An in-house quality assessment of positional reproducibility during RT delivery with vDIBH in patients with left-sided breast cancer was evaluated. The electronic portal imaging device (EPID) was used in cinematographic (CINE) mode to capture a sequence of images during beam delivery. Weekly CINE images were retrospectively assessed for 20 left-sided breast cancer patients receiving RT in vDIBH, and compared with CINE images of 20 patients treated in FB. The intra-beam motion was assessed and the distance from the beam central axis (CA) to the internal chest wall (ICW) was measured on each CINE image. These were then compared to the planned distance on digitally reconstructed radiograph (DRR). The maximum intra-beam motion for any one patient measurement was 0.30cm for vDIBH and 0.20cm for FB. The mean difference between the distance from the CA to ICW on DRR and the equivalent distance on CINE imaging (as treated) was 0.28cm (SD 0.17) for vDIBH patients and 0.25cm (SD 0.14) for FB patients (P=0.458). The measured values were comparable for patients undergoing RT in vDIBH, and for those in FB. This quality assessment showed that using in-room lasers and skin tattoos to independently monitor breath-hold in vDIBH as detected by 'on-treatment' CINE imaging is safe and effective.