Abstract

A micelle-mediated phase separation without added chelating agents to preconcentrate trace levels of aluminium in parenteral solutions as a prior step to its determination by flow injection inductively coupled plasma optical emission spectrometry has been developed. The enrichment step is based on the cloud point extraction of aluminium with the non-ionic surfactant polyethyleneglycolmono- p-nonylphenylether (PONPE 7.5). The chemical variables affecting the sensitivity of the extractive-spectrometric procedure were studied in detail. After optimization, a preconcentration factor of 200 and a %E higher than 99.9 were achieved. The detection limit (DL) value of aluminium for the preconcentration of 50 ml of parenteral solution was 0.25 μg l −1. The calibration graph using the preconcentration system for aluminium was linear with a correlation coefficient of 0.9997 at levels near the DLs up to at least 200 μg l −1. The developed hyphenated assay, which thoroughly satisfies the typical requirements for pharmaceutical control processes, is appropriate to monitor the aluminium concentration in parenteral nutrition.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.