Abstract
Combinations of microbubbles (MBs) and superparamagnetic iron oxide nanoparticles (SPIONs) are used to fabricate dual contrast agents for ultrasound and MRI. This study examines the viscoelastic and oscillation characteristics of two MB types that are manufactured with SPIONs and either anchored chemically on the surface (MBs-chem) or physically embedded (MBs-phys) into a polymer shell. A linearized Church model was employed to simultaneously fit attenuation coefficients and phase velocity spectra that were acquired experimentally. The model predicted lower viscoelastic modulus values, undamped resonance frequencies and total damping ratios for MBs-chem. MBs-chem had a resonance frequency of approximately 13 MHz and a damping ratio of approximately 0.9; thus, MBs-chem can potentially be used as a conventional ultrasound contrast agent with the combined functionality of MRI detection. In contrast, MBs-phys had a resonance frequency and damping of 28 MHz and 1.2, respectively, and requires further modification of clinically available contrast pulse sequences to be visualized.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.