Abstract

The solid-state gamma camera is now commercially available offering the advantages of a compact and portable system, currently used mainly in the cardiac region. We evaluate the ability of the solid-state gamma camera to depict axillary sentinel lymph nodes (SLNs) in breast cancer patients. Preoperative SLN lymphoscintigraphy (LSG) was performed in 19 patients with breast cancer using the solid-state gamma camera. Immediately thereafter, we performed a second LSG using a single detector Anger-type gamma camera, and compared the findings from the two cameras. Concordant results were obtained in 12 (63%) patients with both cameras. In 4 (21%) patients, axillary SLNs were correctly identified only with the solid-state gamma camera. In these patients, the distance between the SLN and the radiopharmaceutical injection site was closer than that of patients who had concordant results (p = 0.001). We can depict correctly axillary SLNs with the solid-state gamma camera in comparison with the Anger-type gamma camera. This technique would be useful for assessing SLNs in breast cancer patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.