Abstract

Background: Light plays an important role in polymerizing a light-cured resin-based composite as a hard restorative material. The present study was conducted to assess the functional efficacy of various light-curing units (LCUs) available to us. Materials and Methods: The present study was conducted in the Department of Endodontics on 60 different LCUs of different manufacturers. Parameters such as type of light-curing unit, the curing protocol, and frequency of bulb replacement were also recorded. Results: Out of 60 LCUs, 26 were quartz–tungsten–halogen (QTH) and 34 were light-emitting diodes (LEDs). The mean intensity of QTH was 486.3 W/cm2 and of LED was 612.1 W/cm2. The difference was significant (P < 0.05). Nineteen QTH and 22 LEDs were satisfactory, while 7 QTH and 12 LEDs were unsatisfactory. The difference was nonsignificant (P > 0.05). Conclusion: The authors found that maximum satisfactory results were seen with QTH type as compared to LED type.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.