Abstract
SummaryRapid and accurate diagnosis of meningitis/encephalitis (M/E) is essential for successful patient outcomes. The FilmArray® meningitis/encephalitis Panel (MEP) is a multiplexed PCR test for simultaneous, rapid detection of pathogens directly from cerebrospinal fluid (CSF) samples. 94 prospectively collected CSF specimens from patients with clinical suspicion of infective M/E underwent testing for 14 pathogens simultaneously, including Escherichia coli, Haemophilus influenzae, Neisseria meningitidis, and Varicella zoster. MEP demonstrated 95% agreement with current PCR methods, resulting in 16 diagnosed cases of M/E. Typically, the FilmArray® MEP results were delivered within approximately one hour, contrasting with current practices taking up to 5.6 days. Given the significant morbidity and mortality associated with delayed diagnosis of central nervous system infections, the FilmArray® MEP is a useful addition to the diagnostic capabilities of a clinical microbiology department.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.