Abstract

The purpose of this clinical study was to assess low frequency, low intensity magnetic fields in the enhancement of the physical rehabilitation of patients after knee endoprosthesis surgery. The study included 62 patients who underwent total knee arthroplasty. Group A consisted of 32 patients who were physically rehabilitated. Group B consisted of 30 patients who were physically rehabilitated and treated additionally with pulsing magnetic fields (5 mT, 30 Hz, 20 min once a day, 5 days weekly). Therapy lasted 3 weeks for both groups. The rehabilitation process was evaluated using a goniometer, tensometer, goniometric pendulum test, Lysholm scale for knee function, and a visual analogue scale (VAS) questionnaire for pain and activity. The changes of measured rates were comparable in both groups. Low frequency and low intensity magnetic fields of examined parameters were not demonstrated to effectively improve the rehabilitation of patients after knee endoprosthesis surgery.