Abstract

We have tested the characteristics of the method of Iwata and Nishikaze (Clin Chem 25: 1317, 1979). The linearity, sensitivity, and precision are satisfactory and the reactivity of benzethonium chloride with various proteins (albumin, immunoglobulins) is the same. The method has been compared with Meulemans's technique (Clin Chim Acta 5: 757, 1960), routinely used in our laboratories, by analysis of 82 samples of cerebrospinal fluid (CSF) and 119 samples of urine. Our results for cerebrospinal fluid agree well with those of Iwata and Nishikaze (r = 0.976; y = 0.992x - 0.013), but we find their method unsuitable for urinary protein determination, probably because of interfering compounds in urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.