Abstract

By Doppler echocardiography, the performance of heart valve prostheses is assessed with the aid of maximal transprosthetic velocities, which, however, may not be representative for the full spatial velocity profile in the vicinity of mechanical valve substitutes due to flow separation by the open occluder. The purpose of this study was to determine characteristics of velocity profiles downstream of a normally functioning Björk-Shiley prosthesis. In a pulsatile flow apparatus, different flow rates of 6.3 and 8.4 l/min were delivered. Using a spatially and temporally resolving ultrasonic Doppler method, velocity profiles 20 and 30 mm distal from the prosthesis were registered and displayed in a three-dimensional grid. The spatial velocity profile was found to deviate substantially from a flat profile at these transducer positions at the two flow conditions. Distal to the minor orifice, velocities measured only 70 and 80% of those downstream of the major orifice. In between, a region of relatively slow moving flow was present. The shape of the profiles remained essentially unchanged during acceleration and deceleration of flow. Thus, spatially resolved velocity profiles downstream of mechanical prostheses can be registered by an ultrasonic Doppler device. These findings may be useful for the detection of beginning malfunction both in the experimental and the clinical setting.