Abstract

The aim of this study was to assess the accuracy of sequential same arm blood pressure measurement by the mercury sphygmomanometer with the oscillometric blood pressure measurements from a device that also determines arterial elasticity. A prospective, multicentre, clinical study evaluated sequential same arm blood pressure measurements, using a mercury sphygmomanometer (Baumanometer, W. A. Baum Co., Inc., Copiague, New York, USA) and an oscillometric non-invasive device that calculates arterial elasticity (CVProfilor DO-2020 Cardiovascular Profiling System, Hypertension Diagnostics, Inc., Eagan, Minnesota, USA). Blood pressure was measured supine in triplicate, 3 min apart in a randomized sequence after a period of rest. The study population of 230 normotensive and hypertensive subjects included 57% females, 51% Caucasians, and 33% African Americans. The mean difference between test methods of systolic blood pressure, diastolic blood pressure, and heart rate was -3.2 +/- 6.9 mmHg, +0.8 +/- 5.9 mmHg, and +1.0 +/- 5.7 beats/minute. For systolic and diastolic blood pressure, 60.9 and 70.4% of sequential measurements by each method were within +/- 5 mmHg. Few or no points fell beyond the mean +/- 2 standard deviations lines for each cuff bladder size. Sequential same arm measurements of the CVProfilor DO-2020 Cardiovascular Profiling System measures blood pressure by an oscillometric method (dynamic linear deflation) with reasonable agreement with a mercury sphygmomanometer.