Abstract

The Ethos treatment planning system allows for the rapid generation of online adaptive treatment plans while the patient is on the treatment couch. One promising application of online adaptive radiotherapy is its use in stereotactic radiotherapy. The purpose of this study was to ensure the Ethos treatment planning system (TPS) can produce clinically acceptable stereotactic plans, that are non-inferior to those from the Eclipse TPS. Forty patients that received previous stereotactic radiotherapy treatment on a Halcyon, 20 of which were lung cases, and 20 that were brain cases, were replanned using the Ethos TPS. The generated IMRT and VMAT plans were compared to the clinical Eclipse VMAT plan. This study found that the Ethos TPS can produce VMAT plans of equivalent quality (target coverage, conformity and OAR doses) to those from the Eclipse TPS for lung SBRT and brain SRT. The IMRT plans produced by the Ethos planning system were marginally inferior to Eclipse VMAT plans, with the differences likely primarily due to beam geometry rather than the optimization system. Ethos plans were generally more modulated than Eclipse plans. With careful selection of optimization structures and reduction in the body contour, VMAT plan generation time could be reduced by 87%. Ethos can generate stereotactic VMAT plans that are equivalent to those from Eclipse in the timeframe required for online adaptive radiotherapy.