Abstract

Cochlear implants (CI) are electronic devices used to restore partial hearing to people with severe hearing impairment. This paper aims to investigate if the introduction of a CI has an effect on SAR distribution in a head model exposed to radiofrequency electromagnetic fields (RF EMF) at mobile communication frequencies. The head model was obtained by image segmentation, the implant was modelled as a geometric structure and the exposure source was modelled as a uniform plane wave at 900 MHz, 1750 MHz and 1950 MHz, incident on the side of the head with the CI. Vertical and horizontal polarizations were simulated. Results show that the presence of a CI inside the cochlea produces negligible variations in the averaged SAR values, both in the head and in the cochlear tissues, although very localized differences in point SAR were found in the cochlea. Globally, these results suggest that finding harmful effects in the cochlear tissues will be unlikely.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.