Abstract
High density Utah slanted electrode arrays (HD-USEAs) have been developed recently for intrafascicular access to submillimeter neural structures. Insertion of such high electrode density devices may cause nerve crush injury, counteracting the intended improved selective nerve fiber access. HD-USEAs were implanted into sciatic nerves of anesthetized rats. Nerve function was assessed before and after HD-USEA implantation by measuring changes in evoked muscle and nerve compound action potentials and single unit neuronal recordings. Neural activity was recorded with over half of all implanted electrodes. Average decreases of 38%, 36%, and 13% in nerve, medial gastrocnemius, and tibialis anterior compound action potential amplitudes, respectively, were observed following array implantation. Only 1 of 8 implantations resulted in loss of all signals. These studies demonstrate that HD-USEAs provide a useful neural interface without causing a nerve crush injury that would otherwise negate their use in acute preparations (<12 h).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.