Abstract

Plasma coagulation as a form of electrocautery is used in liver surgery for decades to seal the large liver cut surface after major hepatectomy to prevent hemorrhages at a later stage. The exact effects of plasma coagulation on liver tissue are only poorly examined. In our porcine model, the coagulation effects can be examined close to the clinical application. A combined laser Doppler flowmeter and spectrophotometer documents microcirculation changes during coagulation at 8 mm tissue depth noninvasively, providing quantifiable information about hemostasis beyond the subjective clinical impression. The temperature at coagulation site is assessed with an infrared thermometer prior and post coagulation and with a thermographic camera during coagulation, a measurement of the gas beam temperature is not possible due to the upper threshold of the devices. The depth of coagulation is measured microscopically on hematoxylin/eosin stained sections after calibration with an object micrometer and gives an exact information about the power setting-coagulation depth-relation. The sealing effect is examined on the bile ducts as it is not possible for a plasma coagulator to seal larger blood vessels. Burst pressure experiments are carried out on explanted organs to rule out blood pressure related effects.