Abstract

During their service life, modular interfaces experience tribological, and corrosion phenomena that lead to deterioration, which in turn can cause a revision procedure to remove the failed prosthesis. To achieve a clearer understanding of the surface performance of those biomedical alloys and the role of the surface properties in the mechanical and chemical performance, samples were taken from retrieval implants made of Ti6Al4V and Co28Cr6Mo alloys. Polarization resistance and pin-on-disk tests were performed on these samples. Physical properties such as contact angle, roughness, microhardness, and Young's modulus were determined. A correlation between surface energy and evolution of the tribological contact was observed for both biomedical alloys. In tribocorrosion tests, titanium particles seem to remain in the surface, unlike what is observed in CoCr alloys. These metallic or oxidized particles could cause necrosis or adverse tissue reactions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.