Abstract

This work reports on the use of two different Monte Carlo codes (GEANT4 and MCNPX) for assessing the dose reduction using bismuth shields in computer tomography (CT) procedures in order to protect radiosensitive organs such as eye lens, thyroid and breast. Measurements were performed using head and body PMMA phantoms and an ionisation chamber placed in five different positions of the phantom. Simulations were performed to estimate Computed Tomography Dose Index values using GEANT4 and MCNPX. The relative differences between measurements and simulations were <10 %. The dose reduction arising from the use of bismuth shielding ranges from 2 to 45 %, depending on the position of the bismuth shield. The percentage of dose reduction was more significant for the area covered by the bismuth shielding (36 % for eye lens, 39 % for thyroid and 45 % for breast shields).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.