Abstract

ObjectivesPoint of care testing (POCT) offers the possibility of near bedside patient testing with a reduction of the turn-around time of analysis. The aim of our study was to determine the analytical performances and usability of a recently developed POCT device for the measurement of tests related to liver function. We evaluated the performance of a liver tests panel performed on the LINX EVO® POCT device. Design and methodsThe imprecision was determined with the Bio-Rad Liquichek Unassayed Chemistry Control. Method comparison was performed with a Cobas® 8000 analyzer. Samples from twenty healthy volunteers were used to verify the reference intervals. Furthermore, practicality was assessed by the healthcare staff handling the POCT device through a dedicated questionnaire. ResultsThe imprecision observed was matching the criteria for the in-lab assay with only one exception, globulin, with an observed imprecision of 6.3 % and a criteria of 5.7 %. With the exception of total and direct bilirubin, the POCT method showed good agreement with the in-lab methods. The verification of reference intervals showed that more than 90 % of the healthy volunteer values were included into the reference interval claimed by the manufacturer except for glucose and globulin. The POCT practicality questionnaire was satisfying overall for users. ConclusionsOur study showed very good analytical performances overall for the liver test panel performed on the LINX EVO® POCT instrument.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call