Abstract

As a phosphorus-containing molecule, nicotinamide adenine dinucleotide is visible by phosphorus magnetic resonance spectroscopy (31P-MRS). However, the relatively low cellular levels of its oxidised (NAD+) and reduced (NADH) forms and a significant peak overlap hinder their evaluation in live tissues. This problem is critical when using 31P-MR spectroscopic imaging, where signals are localised from limited tissue volumes. We have reported improvements in spectral resolution of 31P-MRSIof human tissues in situusing a strict optimisation of the static magnetic field (B0 shimming) and 1H-irradiation during 31P acquisition. Given this,we aimed to demonstrate if these improvements allowed us to measure the in vivo intracellular levels of NAD+ and NADH at the relatively low magnetic field of 1.5 tesla(T). Our results show the feasibility of the in vivo determination of NAD+ and NADH from relatively small volumes of human tissues studied at 1.5 T. These results are clinically relevant as the currently available systems for human use mainly operate at 1.5 or 3.0.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call