Abstract

PurposeA need remains for alternative devices for aerosol drug delivery that are low cost, convenient and easy to use for the patient, but also capable of producing small-sized aerosol particles. This study investigated the potential of recent high power electronic nicotine delivery systems (ENDS) as aerosol generation devices for inhaled bronchodilators. MethodsThe particle size distribution was measured using a cascade impactor. The delivery of terbutaline sulfate, a current bronchodilator used for asthma or COPD therapy by inhalation, was studied. This drug was quantified by liquid chromatography coupled with tandem mass spectrometry. ResultsThe particle size distribution in terms of mass frequency (in two ways, gravimetrically and quantitatively through drug assay on each stage) and the terbutaline sulfate concentration in the aerosol were elucidated. The mass median aerodynamic diameter (MMAD) and the drug delivery rose when the power level increased, to reach 5.6±0.4μg/puff with a MMAD of 0.78±0.03μm at 25W. ConclusionNew generation high-power ENDS are very efficient to generate carrier-droplets in the submicron range containing drug molecules with a constant drug concentration whatever the size-fractions. ENDS appear to be highly patient-adaptive.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.