Abstract

The first derivative of the pseudo-absorption spectrum of a water-loaded cotton wool (water-CW) phantom, which mimics muscle tissues, was used to determine the light path length in the near-infrared (NIR) region. The light path length increased as the density of the turbid medium decreased. It is independent of both water content in the range of 75-85% (by weight) and the diffuse reflecting reference used to determine the pseudo-absorbance. The path length determination procedure was verified by measurements of diffuse reflectance in chicken breast tissue for which the path length of 1.8 mm (differential path length factor, DPF = 2.1) was found to be similar to the path length of NIR light of 1.5-2.2 mm (DPF = 1.8-2.6) in a water-CW phantom of density similar to chicken breast. We conclude that the NIR light path length can serve as a characteristic of muscle tissue density.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.