Abstract

Abstract The mobile phase flow resistances (φ's) of several stationary phases consisting of surface-modified UnisphereTM fused microplatelet alumina particles were determined and compared with corresponding values obtained from stationary phases based upon more conventional spherical and angular silica and alumina particles. Although the φ value obtained for a fused-microplatelet alumina-based phase was slightly lower than the value obtained on a spherical alumina-based phase with similar particle dimensions and surface modification, factors other than particle shape were found to have a substantial effect on mobile phase flow resistance. Larger φ values were obtained with phases with smaller average particle diameters. These were attributed to the presence of greater numbers of microparticles with diameters less than 5 microns in such phases. Larger φ values obtained for wide pore (21 nm) and octyl-bonded alumina-based phases over narrow pore (11 nm) and octadecyl-bonded phases were attributed to the for...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.