Abstract

This study was an assessment of potential exposures of medical personnel to nitrogen oxides during simulated and actual inhaled nitric oxide treatment of newborn and pediatric patients. Breathing zone exposures to nitric oxide (NO) and nitrogen dioxide (NO(2)) were monitored using data-logging personal dosimeters during simulated and actual administration of NO gas to patients in an intensive care setting. Sample. A total of 28 bedside nurses and 18 respiratory therapists were monitored during 6 different patient treatments. The highest measured concentrations of NO and NO(2) in the personal breathing zones of the nurses and respiratory therapists were peak readings (<1 minute in duration) of 6.7 parts per million (ppm) NO and 3.1 ppm NO(2). Exposures averaged throughout 15 minutes and throughout the work shift were below the limit of detection (0.8-ppm NO and 0.5-ppm NO(2)). Detectable exposures to NO and NO(2) were brief, infrequent, and well below Occupational Safety and Health Administration permissible exposure limits or any other exposure guideline, eg, American Conference of Governmental Hygienists Threshold Limit Values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.