Abstract

Single transpulmonary thermodilution (SD) with extravascular lung water index (EVLWI) could become a new tool to better assess lung graft edema during ex-vivo lung perfusion (EVLP). In this study we compare EVLWI with conventional methods to better select lungs during EVLP and to predict post-transplant primary graft dysfunction (PGD). We measured EVLWI, arterial oxygen/fraction of inspired oxygen (P/F) ratio, and static lung compliance (SLC) during EVLP in an observational study. At the end of EVLP, grafts were accepted or rejected according to a standardized protocol blinded to EVLWI results. We compared the respective ability of EVLWI, P/F, and SLC to predict PGD. Mann-Whitney U-test, Fisher's exact test, and receiver-operating characteristic (ROC) curve data were used for analysis. p < 0.05 was considered statistically significant. Thirty-five lungs were evaluated by SD during EVLP. Three lungs were rejected for pulmonary edema. Thirty-two patients were transplanted, 8 patients developed Grade 2 or 3 PGD, and 24 patients developed Grade 0 or 1 PGD. In contrast to P/F ratio, SLC, and pulmonary artery pressure, EVLWI differed between these 2 populations (p < 0.001). The area under the ROC for EVLWI assessing Grade 2 or 3 PGD at the end of EVLP was 0.93. Donor lungs with EVLWI >7.5 ml/kg were more likely associated with a higher incidence of Grade 2 or 3 PGD at Day 3. Increased EVLWI during EVLP was associated with PGD in recipients.