Abstract

To investigate left atrial volumes and function and their variability in healthy volunteers using steady state free precession (SSFP) and fast low angle shot (FLASH) sequences at both 1.5 and 3 T using both the short-axis and biplane area-length methods. Ten healthy volunteers underwent CMR at both 1.5 and 3 Tesla. The biplane area-length method utilized volumes from the horizontal and vertical long axis images. There were no significant differences between left atrial short-axis volumes or function between 1.5 and 3 T assessed using either FLASH or SSFP sequences. The biplane area-length method underestimated maximal left atrial volume using FLASH by 12 mL at 3 T (18%) and by 10 mL (14%) at 1.5 T (p = 0.003 and p = 0.05 respectively). Variability was larger for left atrial measurements using the biplane area-length method. Field strength had no effect on left atrial volume and function assessment using either FLASH or SSFP. The use of the short-axis method for the acquisition of left atrial parameters is more reproducible than the biplane area-length for serial measurements.