Abstract

To investigate the accuracy of intraocular pressure (IOP) as measured by a Reichert Ocular Response Analyzer (ORA), as well as the relationship between central corneal thickness (CCT) and IOP as measured by ORA, Goldmann applanation tonometry (GAT), and dynamic contour tonometry (DCT). A total of 158 healthy individuals (296 eyes) were chosen randomly for measurement of IOP. After CCT was measured using A-ultrasound (A-US), IOP was measured by ORA, GAT, and DCT devices in a randomized order. The IOP values acquired using each of the three tonometries were compared, and the relationship between CCT and IOP values were analyzed separately. Two IOP values, Goldmann-correlated IOP value (IOPg) and corneal-compensated intraocular pressure (IOPcc), were got using ORA. Three groups were defined according to CCT: 1) thin cornea (CCT<520µm); 2) normal-thickness cornea (CCT: 520-580µm); and 3) thick cornea (CCT>580µm) groups. In normal subjects, IOP measurements were 14.95±2.99mmHg with ORA (IOPg), 15.21±2.77mmHg with ORA (IOPcc), 15.22±2.77mmHg with GAT, and 15.49±2.56mmHg with DCT. Mean differences were 0.01±2.29mmHg between IOPcc and GAT (P>0.05) and 0.28±2.20mmHg between IOPcc and DCT (P>0.05). There was a greater correlation between IOPcc and DCT (r=0.946, P=0.000) than that between IOPcc and GAT (r=0.845, P=0.000). DCT had a significant correlation with GAT (r=0.854, P=0.000). GAT was moderately correlated with CCT (r=0.296, P<0.001), while IOPcc showed a weak but significant correlation with CCT (r=-0.155, P=0.007). There was a strong negative correlation between CCT and the difference between IOPcc and GAT(r=-0.803, P=0.000), with every 10µm increase in CCT resulting in an increase in this difference of 0.35mmHg. The thick cornea group (CCT>580µm) showed the least significant correlation between IOPcc and GAT (r=0.859, P=0.000); while the thin cornea group (CCT<520µm) had the most significant correlation between IOPcc and GAT (r=0.926, P=0.000). The correlated differences between IOPcc and DCT were not significant in any of the three groups (P>0.05). Measurement of IOP by ORA has high repeatability and is largely consistent with GAT measurements. Moreover, the ORA measurements are affected only to a small extent by CCT, and are likely to be much closer to the real IOP value than GAT.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call