Abstract

Direct assays for the determination of HDL-cholesterol (HDL-C) have recently become available. The methods are precise, require small sample volume, and appear to be less affected by increased triglycerides than traditional precipitation methods. In this study, we describe the inter- and intralaboratory variability of the Boehringer Mannheim Corporation direct HDL-C assay and its performance in external proficiency testing surveys. A comparison study among three laboratories, using different analyzers and 85 serum specimens, showed a correlation coefficient (r) of 0.99. The direct HDL-C assay also showed good agreement with the ultracentrifugation-dextran sulfate-Mg2+ method (r = 0.98) and the Cholesterol Reference Method Laboratory Network-Designated Comparison Method (a = 0.98x + 4.75 mg/L, r = 0.98). Total error at medical decision levels ranged from -0.8% to +11.1%. Furthermore, this assay performed adequately in the College of American Pathologists and the ALERT surveys as well as the CDC Lipid Standardization Program and met all performance criteria of regulatory agencies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.