Abstract

The AN69 ST membrane was designed to render the surface of the native polyacrylonitrile polymer less cationic. This was achieved by layering the membrane with the polycationic biopolymer polyethyleneimine. This new membrane is able to bind heparin to its surface, through electrical interactions, without altering the reactivity of the sulfonate groups of the membrane, regularly distributed in the membrane bulk. The kinetics of unfractionated or low-molecular-weight heparins were studied in vitro and in vivo in sheep. Encouraging results were obtained indicating that heparin-coated hemodialyzers are potent anticoagulants. Priming the AN69 ST membrane-equipped hemodialyzer with heparin, as in regular hemodialysis, could allow drastic reduction of heparin consumption in hemodialysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.