Abstract

BackgroundGrip strength measurement is widely used in daily medical practice, and it has been reported that the grip strength decreases in patients with carpal tunnel syndrome (CTS). However, conventional grip dynamometers evaluate only the maximum power of total grip strength and cannot measure the time course of grip motion. In this report, we aimed to determine the grip characteristics of CTS patients by measuring the time course of each finger’s grip motion and to analyze the relationship between finger grip strength and subjective symptoms using this new grip system.MethodsThe grip strength of each finger was measured using the new grip system that has four pressure sensors on the grip parts of each finger of the Smedley grip dynamometer. We analyzed the time course of grip motion and relationship between finger grip strength and subjective symptoms in 104 volunteer and 51 CTS hands. The Japanese Society for Surgery of the Hand version of the Carpal Tunnel Syndrome Instrument (CTSI-JSSH) and the Disability of Arm, Shoulder, and Hand questionnaire (DASH) were used as subjective evaluation scores.ResultsIn the CTS group, the grip time with the index, middle, and ring fingers was longer, and the time at which strength was lost after reaching the maximum was earlier. Patients with severe subjective symptoms tended to not use the index and middle fingers during grip motion.ConclusionsThis new system that measures each finger’s grip strength at one time and record the time course of grip motion could quantify a patient’s symptoms easily and objectively, which may contribute to the evaluation of hand function.

Highlights

  • Grip strength measurement is widely used in daily medical practice, and it has been reported that the grip strength decreases in patients with carpal tunnel syndrome (CTS)

  • To understand the pathophysiology of CTS, we aimed to analyze the grip characteristics of CTS patients by measuring the time course of the grip motion, and we sought to determine the relationship between finger grip strength and subjective symptoms using this new grip system

  • Time course of each finger’s grip strength Grip time with the index, middle, and ring fingers was longer in the CTS group than in the control group (Fig. 2)

Read more

Summary

Introduction

Grip strength measurement is widely used in daily medical practice, and it has been reported that the grip strength decreases in patients with carpal tunnel syndrome (CTS). Conventional grip dynamometers evaluate only the maximum power of total grip strength and cannot measure the time course of grip motion. We aimed to determine the grip characteristics of CTS patients by measuring the time course of each finger’s grip motion and to analyze the relationship between finger grip strength and subjective symptoms using this new grip system. A conventional grip dynamometer, which has been used in many studies that have demonstrated weakened grip strength in patients with CTS, evaluates only the maximum power in the total grip strength of all fingers. It was necessary to develop a grip dynamometer that can measure the time course to clarify the characteristics of grip motion in CTS patients

Objectives
Methods
Results
Conclusion