Abstract

This study provides a quantitative assessment of the accuracy of a commercially available deformable image registration (DIR) algorithm to automatically generate prostate contours and additionally investigates the robustness of radiomic features to differing contours. Twenty-eight prostate cancer patients enrolled on an institutional review board (IRB) approved protocol were selected. Planning CTs (pCTs) were deformably registered to daily cone-beam CTs (CBCTs) to generate prostate contours (auto contours). The prostate contours were also manually drawn by a physician. Quantitative assessment of deformed versus manually drawn prostate contours on daily CBCT images was performed using Dice similarity coefficient (DSC), mean distance-to-agreement (MDA), difference in center-of-mass position (ΔCM) and difference in volume (ΔVol). Radiomic features from 6 classes were extracted from each contour. Lin’s concordance correlation coefficient (CCC) and mean absolute percent difference in radiomic feature-derived data (mean |%Δ|RF) between auto and manual contours were calculated. The mean (± SD) DSC, MDA, ΔCM and ΔVol between the auto and manual prostate contours were 0.90 ± 0.04, 1.81 ± 0.47 mm, 2.17 ± 1.26 mm and 5.1 ± 4.1% respectively. Of the 1,010 fractions under consideration, 94.8% of DIRs were within TG-132 recommended tolerance. 30 radiomic features had a CCC > 0.90 and 21 had a mean |%∆|RF < 5%. Auto-propagation of prostate contours resulted in nearly 95% of DIRs within tolerance recommendations of TG-132, leading to the majority of features being regarded as acceptably robust. The use of auto contours for radiomic feature analysis is promising but must be done with caution.

Highlights

  • This study provides a quantitative assessment of the accuracy of a commercially available deformable image registration (DIR) algorithm to automatically generate prostate contours and investigates the robustness of radiomic features to differing contours

  • Twenty-nine prostate cancer patients who were treated with volumetric modulated arc therapy (VMAT) and had daily cone-beam CTs (CBCTs) images were considered

  • A previous study by Studenski et al, considered 16 patients on hypo-fractionated schemes and found percent difference in prostate volume from contours created through a DIR workflow to be < 10%38, and a previous study by Forde et al, which looked at inter-observer delineation variability and its impacts on radiomic feature robustness found the interobserver variation in contour sizes were as high as 80%5

Read more

Summary

Introduction

This study provides a quantitative assessment of the accuracy of a commercially available deformable image registration (DIR) algorithm to automatically generate prostate contours and investigates the robustness of radiomic features to differing contours. Planning CTs (pCTs) were deformably registered to daily cone-beam CTs (CBCTs) to generate prostate contours (auto contours). Quantitative assessment of deformed versus manually drawn prostate contours on daily CBCT images was performed using Dice similarity coefficient (DSC), mean distance-to-agreement (MDA), difference in center-of-mass position (ΔCM) and difference in volume (ΔVol). Auto-propagation of prostate contours resulted in nearly 95% of DIRs within tolerance recommendations of TG-132, leading to the majority of features being regarded as acceptably robust. This study extracted a total of 137 radiomic features from planning CT images of head and neck cancer patients and warned that variation in delineation can significantly affect some radiomic features. The American Association of Physicists in Medicine (AAPM) Task Group 132 (TG-132) provides recommendations on the use of image registration and fusion algorithms and provides quantitative methods of evaluating DIR a­ ccuracy[20]

Objectives
Methods
Results
Discussion
Conclusion