Abstract

Objectives Lp-PLA2 is a biomarker with promise for predicting cardiac risk. The lack of reproducible results has limited its use. In evaluating a new reagent kit, we investigated conditions for optimal reproducibility. Methods The Auto-PLAC reagents were evaluated on the Cobas instrument. Performance characteristics, stability, and population ranges were determined. Results Analytical performance characteristics replicated manufacturer's claims. The stability profile of the analyte was unusual, with increasing results observed with storage at 4 °C or − 20 °C. Only storage at − 70 °C gave acceptable stability. Population median values with properly preserved samples were much lower than the cut off previously validated for increased risk. Conclusions It is postulated that variability in specimen handling was a major contributor to the lack of traceability of the current reagents to the earlier clinical studies demonstrating its utility. We are now unsure how to identify reliable criteria for result interpretation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.