Abstract
The value of laser Doppler flowmetry in measuring blood flow through the lung was assessed comparing it with flow measured by electromagnetic flowmetry. This was an experimental laboratory-based prospective study performed in an approved University animal research facility. Ten beagle dogs were used. Simultaneous measurement of pulmonary blood flow by laser Doppler flowmetry and electromagnetic flowmetry was carried out at varying degrees of pulmonary artery constriction. There was a linear relationship between the two methods of assessing blood flow (regression equation: y = 0.9x + 5.5; p = .00001) using a least-squares, best fit, straight line analysis (160 data points). The results of this study demonstrate that laser Doppler flowmetry provides an accurate indicator of pulmonary blood flow, which in practice is both simple and reproducible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.