Abstract

The aim of our study was to compare the biomechanical stability of three different systems used to fix mandibular condylar fractures on synthetic polyurethane mandibles using a servohydraulic testing unit and finite element analysis (FEA). We measured the resistance to displacement loads causing 1.75 and 3.5mm displacements of the fragments, and the maximum resistance values before failure of the fixation system in 30 condylar necks of 15 polyurethane mandible models. The three subgroups investigated were arranged as a triangular plate (n=10), a quadrangular plate (n=10), and a straight miniplate (n=10). The most successful fixation pattern was achieved with two straight four-hole miniplates. However, the triangular plate can easily be applied as a second option by an endoscopic approach. The rectangular plates showed no success in mandibular condylar fractures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.