Abstract

ABSTRACT Purpose To assess the repeatability of several anterior segment parameters obtained with swept-source optical coherence tomography (SS-OCT). Methods We measured 5-times consequently several parameters in 69 eyes using the ANTERION SS-OCT. White-to-white (WTW), angle-to-angle (ATA), spur-to-spur (STS), and lens vault distances were measured in the horizontal- and vertical-meridian. Anterior-chamber-angle (ACA), scleral-spur-angle (SSA), angle-opening-distance (AOD), and trabecular-iris-space-area (TISA) were measured at 500 and 750 μm for the superior-, nasal-,inferior-, and temporal-quadrants. Intrasubject standard deviation (Sw),coefficient of variation (CoV), coefficient of repeatability (CoR), and intraclass-correlation-coefficient (ICC) were calculated for each parameter. Bland-Altman analysis was done. Results We have not found statistically significant differences between repeated measurements (p > 0.05). Repeatability was good for the different parameters evaluated. Sw values in distances and areas were low and ranged from 0.01 to 0.07. CoR values showed a similar pattern being larger for those metrics measuring angles. The same happened with CoV values, being very small for WTW,ATA, and STS distances (0.16–0.57%). ICC values for all parameters analyzed were > 0.97. Bland-Altman plots evidenced the narrow limits of agreement for all parameters. Conclusions The ANTERION SS-OCT demonstrated high repeatability measuring different distances, angles, and areas of the anterior segment of the eye.