Abstract
We report the accuracy of positioning of the calibration ball in the process of pre-operative templating for total hip arthroplasty (THA). The ball should be placed in the coronal plane of the hip to provide suitable accuracy. We reviewed 112 post-operative THA radiographs where a calibration ball had been placed. We templated the femoral head size of the implant after calibrating the templating system from the ball. The calibrated femoral head diameter was compared to the known prosthetic head size. A percentage error was calculated. Overall, incorrect placement of the calibration ball resulted in a mean percentage error in templating of 6.8% (median 5.7%; range 0–26%). Such error carries implications with the templating process and may result in incorrect component sizes, leg lengths and offset.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.